There are three typical modes used to purge the dryer of water vapor. Each of the modes has its advantages and disadvantages but all are simple to configure and operate. In each case, the outlet humidity can be changed by adjusting the purge vacuum level and or flow. Typically, in the case of the split sample and reflux modes, higher purge vacuum reduces sample gas outlet humidity and lower purge vacuum increases sample gas outlet humidity.
“Reflux” mode employs a vacuum pump to purge the dryer. In this mode, a portion to 100% of the dry sample gas that is exhausted from the analyzer is returned to the dryer, expanded through a needle valve or other flow restrictor, and used as the purge gas. This requires a tube in addition to the sample gas tube to be connected to the analyzer.
- Purge vacuum range: 20in-hg vacuum (336 millibar abs.) to full vacuum
- Notes:
1. If the analyzer incorporates its own sample gas pump, it is recommended that the outlet be vented to atmosphere as shown. This will ensure that the pressure in the analyzer is not affected in the event the analyzer’s sample pump and the purge vacuum pump are operated at different flow rates. Depending on the purge vacuum pump performance curve, controlling the purge flow rate at a lower level than the analyzer sample flow rate will be desireable and will increase drying performance.
2. If the analyzer does not incorporate its own sample gas pump, the dryer purge vacuum pump may be used to draw the sample gas through the entire circuit. In this case,the atmospheric vent must be closed.
“Split Sample” mode is essentially the same as the “reflex “mode except the sample gas is split into two flows after it exits the dryer. One stream flows to the analyzer and the other stream flows back to the dryer to be used as the dryer purge gas. The main advantages of this mode over the reflux mode are only one tube is needed to run to the analyzer, the drying process is independent of the analyzer, and none of the sample gas needs to be returned to the dryer once it exhausts from the analyzer. The disadvantage is that this creates the need for additional sample flow through the dryer to be used as purge gas. To maintain a similar level of drying performance to that of the reflux mode, there is a need for a somewhat larger vacuum pump that can generate a deeper vacuum. In order to optimize drying performance, the dryer purge flow rate must be minimized as to not unnecessarily increase the total flow rate through the dryer. This also allows the pump to generate purge vacuum that is as deep as possible.
- Purge vacuum range: 25in-hg vacuum (167 millibar abs.) to full vacuum
“Normal” mode is the most basic configuration. There are no moving parts and it is quieter than having a vacuum pump running continually but requires that instrument grade air be available for use as the dryer purge gas. In general, more purge flow reduces sample gas outlet humidity, less increases sample gas outlet humidity.
- Purge air humidity: Oil-less air, Instrument grade air is required, -40° C range dew point or below.
- Purge air flow rate: 2x the sample flow rate. (Additional flow will be of little or no benefit)