Baldwin-Series Model 34C-R Heated Filter Probe

5 views 0

Symptom Check Action
120 VAC heater jacket is not heating Check the resistance between the black and white wires (tied together) and the blue wire. Resistance should be between 100-130 ohms together or from 230-250 ohms for each leg. If the measure is open for heater resistance the fusible link has blown and the jacket should be replaced.
220 VAC heater jacket is not heating Check the resistance between the black and white power wires using an ohmmeter. Resistance should be between 460-480 ohms. If the measure is open for heater resistance, the fusible link has blown and the jacket should be replaced.
Filter plug cannot be removed from filter housing Check “O” rings for damage High particulate loading Replace “O” rings
Clean the “O” ring sealing surfaces with a clean towel prior to reassembly.


Download Baldwin-Series Model 34C-R Heated Filter Probe User’s Manual