Symptom Diagnostic Fix
120 VAC heater jacket is not heating CAUTION – Disconnect power from the system. Check the resistance between the neutral and hot terminal block. Resistance should be about 30-35 ohms If the measurement is out of range for heater resistance, one or both of the heater cartridges are damaged. Both should be replaced.
220 VAC heater jacket is not heating CAUTION – Disconnect power from the system. Check the resistance between the neutral and hot terminal block. Resistance should be about 120-130 ohms. If the measurement is out of range for heater resistance, one or both of the heater cartridges are damaged. Both should be replaced.


Download Baldwin-Series Model 45 Dilution Probe User’s Manual

Was this helpful?