Filter Replacement:
To determine if filter needs to be replaced, monitor flow rate of sample gas of pump. If output begins to drop, it may indicate a clogged filter element.

  1. Note orientation of filter. Flow direction is upward.
  2. Push lock ring on black elbow fitting twards body of the fitting. This will retract the internal gripper that holds the filter in the fitting.
  3. Pull tube stubs from fitting and repeat for the other end
  4. Replace with new filter (part number: AG-DIF-K70).